Intended use and precautions of carryx-II

1.Overview

1.1 Intended use of this product.

Used for X-ray photography of teeth, to obtain images for clinical diagnosis.

1.2 Composition

This equipment is mainly composed of X-ray tube head, control system, battery, charger and beam limiter.

1.3 Classification
Anti-shock type: Class 1 when charging, internal power supply during operation.
Degree of anti-shock: Type B application part
Non-AP 1 APG type equipment
Degree of anti-liquid: IPXO (closed type without anti-liquid)
Operation mode: intermittent load continuous operation

1.4 User's responsibility
The user must:
*Use this equipment in accordance with the relevant operation and maintenance instructions in this manual. The manufacturer and its agents assume no responsibility for personal injury or property damage, equipment damage, or malfunctions caused by improper use or maintenance of the user.
* In the event of an accident related to this equipment or a change in function or performance that could cause death of a patient or operator, personal injury or health hazard, the health authorities and the manufacturer or their agents should be notified immediately.
* The product model and factory number should be included in the problem feedback report provided by the user to the manufacturer, and the user can obtain this information from the technical label.

1.5 Warnings and precautions
The use and operation position of public equipment is very important. The radiation safety requirements of each country and region are different, and the equipment should be shielded. It is the responsibility of the user to ensure compliance with local safety requirements.
This x-ray machine may produce ionizing radiation, which can be hazardous to health if not properly controlled. Therefore, it is recommended that the machine be operated only by trained personnel in accordance with existing laws and regulations.
Although this unit complies with the electromagnetic compatibility specifications, we also recommend not to use this unit in the place where there is external electromagnetic interference. (Such as high-power equipment or motor equipment, etc., which can interfere with the electronic circuits of this system)
Public users are not allowed to open the casing of the equipment without permission. If the equipment fails, please contact the manufacturer or qualified service personnel to avoid the danger of high voltage electric shock.
Designed replaceable parts, such as batteries, chargers, handheld exposure devices and other parts must use original accessories, users can not replace them at will.

Distributor maintenance personnel authorized by the company can provide drawings and component information.

1.6 lable instruction

NO.lconDescriptionNO.lconDescription
1Type B equipment6fragile
2Note, refer to the
attached file
7No water Resistant
3x-ray focus position8Manufacturing date
4Power switch9Serial number
5lonizing radiation10Exposure Button

2. Technical data

2.1 Technical specification
Charger input:100 - 240V~
Frequency:50/60Hz Power:30VA
Internal power:10.8V
Power resistance:<2Q
Ray type: X ray
Ray direction and distribution: diameter of beam limiter exit direction 60mm
Dose rate:2mGy/s
X ray tube model:D-045
Target material: Tungsten
Target angle:12.5
X Ray focus:0.4mm
Tube voltage:70kV+10%
Tube current:2mAt20%
Loading time adjustment range:0.04s-2.0s deviationt (10%+ 1ms), Adjusting way
please press R'10 Coefficient choice.
Nominal electric power:0. 14kW(70 kV. 2mA.0.1s)
Combination of loading factors that cause maximum output electric power:70 kV. 2mA
Inherent filtration:1 .0mmAl/70kV
Additional filtration:0. .5mmAl/70kV
Total filtration:1 .5mmAl/70kV
Half-value layer:70kV > 1.6 mmAl
Load cycle:1/1 5 Minimum 6 seconds
Radiation leakage rate:1 meter position<0.25mGy/h
Total weight:1.7kg

2.2 Beam limiter
Focus-skin distance:20 .5cmt0. .5cm
Output radiation field: round, ф 6cm:0.5cm

2.3 Controlling system
Microprocessor control
Children(small); Adult(big),Teeth type options

2.4 Working environment requirement
Working temperature:10"C-40"C
Relative humidity:30%RH~ 75%RH
Atmospheric pressure:700hPa~ 1060hPa

2.5 Transportation and stock conditions
Stock temperature:- 20"C-55"C
Transportation temperature: 20C-55"C
Relative humidity:<93%RH
Atmospheric pressure:700hPa~ 1060hPa
Caution: dental x ray unit need to be stocked in the indoor environment without
corrosive gas and with good ventilation.

3.Check list--before using the equipment
Please read the user manual carefully before using, and make sure familiar with
radiation protection measures
Make sure the films meet the working requirement and ready to be used
Make sure the films match with develop liquid
Make sure validity of develop liquid,and put it under proper processing temperature
and concentration
Make sure validity of films, and don't use the expired films
If use other imaging items,please make sure the imaging items are in normal
available status
Make sure the equipment is fully charged

4.Product Composition

1. Beam limiter 2.Power Switch 3.Hand Exposure device 4.Display 5.Function keys button 6.Charge port 7. Battery House 8. Exposer button 9. Support interface

5.Shooting X Ray & Equipment Operation
5.1 Turn on
 
a. Preparation: Insert the battery into the device.
b. Turn On: Long Press the Power ON switch and the LCD display will glow,notifying with beep sound.
c. Check: Check for the sufficient battery level.
d, Mode Selection: Adult/children & then tooth selection.
e. Exposure time: The system has a default time setting, users can also adjust the shooting time as needed
f. Imaging: Choose film or sensor imaging for the patient's teeth in the required area, press the exposure button, emit rays and have a notify sound
g. Cooling: after the exposure, the equipment enters the cooling countdown, standby, ready to expose again
h. Power off the equipment: Long Press the power switch,it shutdown with double beep sound

5.2 Display interface symbols and its meaning

SymbolsDetailsSymbolsDetails
Standby and ready for standbyLow power
ReadyBattery power display
Under exposureCharging
Exposure time is abnormal,
warning symbol
Lock
High temperature warningAdult mode
High pressure warningChildren mode
Upper teeth foretoothlower tooth foretooth
Upper tooth molarlower tooth molar
Upper teeth molarlower tooth molar
bitewingWisdom tooth
Expose timeModeMode select
UpIncrease exposure setup timeTeethReduce exposure setup time 
DownReduce exposure setup time

5.3 Patient position
Keep the patient in the correct sitting position, place film or digital imaging sensor, with the protection. Then adjust the angle and position between the X-ray generator and the patient.

The film or digital sensor placed in the mouth or on a film holder, and the film or sensor should be parallel to the long axis of the tooth.

Ask the patient to hold the film or digital sensor with their fingers on the exposure position in the mouth. Position the X-ray perpendicular to an imaginary line that bisects the angle between the film or digital sensor plane and the long axis of the tooth perpendicularly.

X-ray inclination average angle when shooting upper and lower teeth

Tooth

X-ray tilt direction X-ray tilt direction X-ray tube tilt angle
Maxillary incisorTilt to the foot+42°
Maxillary uni-cuspidTilt to the foot+45°
Maxillary bicuspids and first molarsTilt to the foot+30°
Maxillary second and third molarsTilt to the foot+28°
Mandibular incisorTilt to the head-15°
Mandibular uni-cuspidTilt to the head-18°~-20°
Mandibular bicuspids and first molarsTilt to the head-10°
Mandibular second and third molarsTilt to the head-5°

5.4 exposure
1. Press the exposure button on the device first, wait for the 60-second countdown on the display, and enter the exposure-ready state.
2. At the end of the exposure, the screen will display the dose value of the exposure.At the same time the icon flashes,which indicates that the equipment is cooling.Wait until the icon is closed to prepare for the next exposure
Attention:lf it is the first use of the new machine,Please adjust the exposure time to the value of 0.1s for the exposure test,when the machine is properly exposed, select another value to expose.

5.5 Charging and battery maintenance
5.5.1 Charging
al Turn on the power switch of the device;
b/ Connect one end of the charger to the device's charging port and the other end to
the network power (100-240V ~ 50 1 60Hz);
c/A charging icon appears on the screen when the device is connected to the charger
d 1 When the battery level is full, means charging complete.
e 1 Disconnect the charger, once the charging completed.
f/ Full Charge: about 4 hours.
Tip: you can turn off the device while charging

5.5.2 Battery Maintenance
a 1 When the machine is not in use, the power switch should be turned off to save power.
b/ Please use the original charger to charge.
c/ When the battery is not used for a long time, the battery should be separated from the equipment, and it should also be charged once in a three months.
d /lt is better to keep it atleast 20% charge.
e 1 Should avoid high and long-term single charge for more than 12 hours.

5.6 Lock device exposure
In order to prevent misoperation,this device have lock function
Lock: when the Mode key is pressed and the Teeth key is pressed, the upper right side of the display screen will be displayed 8 , indicating that the device's exposure function is locked. At this time, other functions operate,the device cannot be exposed by pressing the exposure key.
Unlock: when you press the Mode key and the Teeth key again, the ICONS 6 on the upper right of the display close, indicating that the lock is released and the device can be exposed normally

5.7 View the cumulative dose
The device has the function of cumulative exposure dose
When the Mode key is pressed, and then pressed the UP key, the screen shows the cumulative dose of the device that has been exposed, and the unit is uGy or mGy or kGy, When the Mode and UP keys are pressed again to exit the dose to see the status.

5.8 Exposure time setting
When the default exposure time of the device does not meet the user's current imaging requirements, the user can setup the time of tooth position. 
Setting method: Select the tooth position, adjust the exposure time by UP or DN button. Press UP and DN button together for 3 seconds after adjusting the required exposure time, the system will sound "DiDi"to indicate that the current exposure time has been saved.
Restore device's default exposure time: If the user wants to restore the exposure time of the device to the factory state, please press UP and DN button together for 10 seconds, the device will emit a long "Di' sound. The system will return to the factory default exposure time.

6.wrong instructions

Error
indication
Error infoSolution
Device exceptionRe-expose or view the exposure code
High temperature warningNeed to be cooled for 0.5-1 hours
High voltage warningRestart, if the fault persists,please contact the manufacturer
Low batteryPlease charge on time
Error
indication
Error infoSolution
E01Exposure terminatedPress and hold the exposure button
until the exposure finish
E02KeepRestarting the device to expose,
If still fails, please contact the dealer
E03High voltage warningRestarting the device to expose,
If still fails, please contact the dealer
E04The voltage of the X-ray
tube is too high
Restarting the device to expose,
If still fails, please contact the dealer
E05The current of the X-ray
tube is too low
Restarting the device to expose,
If still fails, please contact the dealer
E06The voltage of the X-ray
tube is too low
Restarting the device to expose,
If still fails, please contact the dealer
E07Anode current feedback failure

Check the connection between the control panel and the ball tube or contact the dealer

E08Anode voltage feedback failureCheck the connection between the control
panel and the ball tube or contact the dealer

7. Security Features
. Only trained and authorized technicians are allowed to open the case and touch the circuit board.
. The equipment must be switched off before cleaning or disinfecting the equipment.
. Water and other liquids must not penetrate the interior of the device, as they may cause short circuits or corrosion.
. This equipment cannot be used in a flammable gas or steam environment.
. Only trained and qualified personnel should be allowed to operate this equipment and must comply with existing regulations regarding radiation protection.
. Patients and operators should wear protective gear when taking X-rays.

. Make sure that this equipment is qualified.
. This machine contains parts that must be disposed of in accordance with existing regulations.
. Contraindications: Use with caution in pregnancy cases.

8. Cleaning and disinfection
8.1 Cleaning

The power must be disconnected before cleaning. Use mild instrument disinfectant carefully to remove fingerprints and other stains. Do not allow liquids to enter the device. You can use a soft cloth to clean the plastic case.
8.2 Disinfection
Parts that are accidentally touched by the patient must be cleaned with a cleaning agent (eg 2% ammonia) and then disinfected. Do not use solvents or corrosive disinfectants.

9. Maintenance and inspection
9.1 Daily maintenance

Make sure that the device is placed where only the operator can get it;  Please reserve a clean cloth to regularly clean the device.
9.2 Inspection
a/Daily inspection
-The boot check screen can be displayed normally to check whether the keys can be used normally
-Check if the buzzer sound works Ok.
-Check if the exposure key and indicator are normal.
b/Monthly Check
-Please confirm whether the battery life of the used battery pack is normal
c/ Check Every Year
-The equipment must be inspected for safety every year

10. Waste disposal Process
In order to reduce the burden on the environment, when the equipment is scrapped, please dispose of it according to local laws, regulations and environmental requirements. When disposing of waste products, necessary precautions must be taken to protect against damage.