XR-11 operation method and precautions

Lithium battery maintenance

Please make sure charge fully before the first using.

When the machine indicates"E01"must charging immediately. 

if not using for a long time Should charging 80% electricity and 

rechargeable every 3 months to avoid the battery over discharge to cause damage.

The indicate light of the charger from red turn green is mean the battery is fully.

The lithium battery is not resistant to impact, high temperature,damp,

electromagnetic wave, high pressure.

please pay attention to the  environment and gently during the transportation and the using.

Please use the charger with the machine self and do not use inferior or other types of battery charger.

 

Error messages

If there is any error message occurs on the device, the message as follows may help you identifying the problema.

Eo1:May lower battery, the battery needs to charge.

E10: May no filament voltage  

E11:May lower filament voltage

E12:May higher filament voltage 

E13:Filament may always bright

E30:kV may abnormal

E31:kV may lower

E32:kV may higher

E40: mA may abnormal

E41:mA may lower

E42:mA may higher

E50:X-ray generator may overheated

If display E10~E42,turn off the device immediately and use the device again about 5mins later.

if display E50,turn off the device immediately and use the device again about 30mins later.

 

Notice

Once charging had finished(indicator of charger changes form red to green),

disconnect the DC output connect form the charger and pull out the power cord and store it.

Battery is a expendable supply and duration of a battery is shortened slowly after the first purchase.

Please buy a new battery if duration decreases by a half when compared to the first use.

 

Maintenance and keeping

Keep the device in the place where only authorized personal can reach.

Please clean the device daily by cleaning clothes and must leep cone clean.

Please do not remodel the device arbitrary.

Rrecommend to check the device once a year at the designated A/S center for safe use.

Please recharge the battery every three months when it not use for long time,and keep it in store.

 

Daily tests and annual maintenance.

When turn on,piease confirm error message is not display on the LED screen.

Please confirm whether buzzer sound is heard and"EXP" words on the LCD,

and exposure LED could see normally at x-ray exposure.

Every year for safe use,safety check-up is necessary.

 

Other warning

If any problem or error founded in the device, stop operating and then consult with the designated A/S center.

Do not disassemble or modify the device, since it include generates x-ray,if not use properly,it may damage users or patients.

Do not allow an unauthorizd person to use or repair it.

The range of permissible voltage AC100-240V,please use the charger within this range.

There is a potential danger for an electrical shock if you connect the power plug to consent or remove it.

Do not touch the device with your wet hand or make a physical contact with the conducting part.

If  use an unauthorized charger,battery may be damaged.

If not be used for long time,need to charge the unit per month.

Used battery should not be thrown away in trash,please put that in a recycling box or exclusive case.

 

Warning about using

Make sure the person who using this device is legally.

If pregnant, ask a health professional before taking x-ray.

An excessive exposure of raditation leaking may cause human body reaction lightly.

Humidity conditions recommended for storage:10-85%RH.

Humidity conditions recommended for use:15-80%RH.

Optimal humidity range:15-60%RH.

Temperature conditions recommended for storage:10-50℃ 

Temperature conditions recommended for use: 10-35℃.   

Optimal temperature  10-25℃